ACCEPTANCE TESTING
5-26
RF spectrum analyzer
Specification
Frequency range:
400 kHz to 1.0 GHz
Resolution bandwidth
300 Hz, 3, 30, 300 kHz, 3 MHz
Display dynamic range:
80 dB
Noise floor
Typically 75 dB below top of screen
On screen linearity:
Typically ± 2 dB ± resolution (10 dB/div) (10 dB above the noise
floor)
Resolution:
0.1 dB on 2 dB/ division
0.5 dB on 10 dB/ division
Level flatness
±2 dB ± resolution (10 dB/division)
Test equipment
Description Minimum specification Example
RF signal generator
100 kHz to 1 GHz. Level accuracy ±0.85 dB
IFR 2041
RF power meter
±0.1 dB from 10 MHz to 1 GHz
IFR 6960/A/B +
6912 sensor
C5898
N TYPE
UUT
Signal generator
RF OUTPUT
Fig. 5-16 RF spectrum analyzer checks
(1) Refer to Results table 5-32 on page 5-41 . Set the UUT to [SPEC ANA], Vid Filter ON,
10 dB
/
div, Res BW 300 Hz, and Peak Hold OFF, [Centre Freq] 101 MHz, [Span] 1 kHz,
BNC antenna input socket, [Ref Level] -30.0 dBm.
(2) With nothing connected to the UUT, check that the noise floor is within the bottom 10 dB
graticule division of the display. (Between 70 and 80 dBm dynamic range.)
(3) Connect the equipment as shown in Fig. 5-16, connecting the incoming signal to the N-type
input socket of the UUT. Set the UUT to N-type input.
(4) On the UUT, select [Res BW] [Manual Res BW], [30 kHz], [Ref Level] to 0 dBm.
(5) Set the RF signal generator to provide a signal at 101 MHz at a level of 0 dBm with no
modulation. Adjust the RF signal generator level until the trace sits at the top of the screen.
Check that the level set on the RF signal generator is 0 dBm ± 4 dB (Functional test only).
Note this reading (reading 1).
(6) Reduce the RF signal generator RF level until the trace sits on the graticule line 1 division
down from the top line. Check that the level set on the RF signal generator is 10 dB down
from reading 1 ± 3 dB (functional test only).